Product Name:Reverse Osmosis Water System For Hospital
Product Profile:
The product is used for multi bed hemodialysis and related treatment water in medical institutions. Such as for concentration of power preparation ,hemodialysis ,reusing dialyzers.
System Feature:
1. 10” Touch Screen ,exactly control
2. Adopt imported PLC Logic control
3. With automatic protection of low & high pressure
4. With automatic switch feature for the first & second stage to ensure usage of emergency situation
5. Alarm of exceeding standard for water quality of second stage .
6. With manual & automatic can switch when the system is working .
7. Automatically flush circulation loop after stopping to prevent growth of bacteria.
8. Supply water of Circulation loop with constant voltage
9. The RO Part adopt sanitary material UPVC /304
The Parameter:
Type & Spec. | CY-BYXT-Ⅰ-250/500/1000/1250/1500/2000/3000L/H |
The Raw Water Quality | Tap water ,TDS≤200ppm,Pressure 0.2~0.4Mpa,Water Temperature:5~40℃ |
Supply Power | 380VAC、50HZ | Operation Pressure | 0.2-0.6Mpa |
Water Quality | Meet YY0572-2005,meet USA AAMi/ASAIO Standard of hemodialysis water ,removal rate of bacterial endotoxin ≥99%,No dead cavity. |
Production Water Quantity | 250L/H | 500L/H | 1000L/H | 1250L/H | 1500L/H | 2000L/H | 3000L/H |
Power | 2000VA | 2500VA | 3500VA | 4000VA | 5000VA | 7000VA | 10000VA |
Inlet Diameter | DN 20 | DN 25 | DN 32 | DN 40 | DN 40 | DN 40 | DN 50 |
Outlet Diameter | DN 20 | DN 20 | DN 25 | DN 25 | DN 25 | DN 32 | DN 32 |
Suitable Beds | 1-5 | 5-10 | 10-20 | 20-25 | 20-30 | 30-40 | 40-60 |
Size | 2000×750×1600mm | 2200×750×1600mm | 2500×850×1600mm |
Weight | 450kg | 580kg | 800kg | 900kg | 980kg | 1100kg | 1500kg |
Installation Area | 8㎡ | 10㎡ | 15㎡ | 20㎡ | 25㎡ | 35㎡ | 40㎡ |
Note: 1. The temperature requirement of usage &installation is 5-40℃,Humidity ≤80%, There is no direct sunlight in the equipment room , The floor drain is set on the ground and the drainage is sooth .
2.The bearing capacity of room should meet installation of requirement .This product should avoid running in environment with strong electromagnetic interference .